JULABO SE-Z Heating Circulators User Manual
Page 74
Remote control
74
Command
Parameter
Response of circulator
in_par_04
none
CoSpeed - Band limit (max. difference between the
temperatures in the internal bath and external system).
in_par_05
none
Factor pk/ph0:
Ratio of max. cooling capacity versus max. heating capacity
in_par_06
none
Xp control parameter of the internal controller.
in_par_07
none
Tn control parameter of the internal controller.
in_par_08
none
Tv control parameter of the internal controller.
in_par_09
none
Xp control parameter of the cascade controller.
in_par_10
none
Proportional portion of the cascade controller.
in_par_11
none
Tn control parameter of the cascade controller.
in_par_12
none
Tv control parameter of the cascade controller.
in_par_13
none
Adjusted maximum internal temperature of the cascade
controller.
in_par_14
none
Adjusted minimum internal temperature of the cascade
controller.
in_par_15
none
Band limit (upper)
in_par_16
none
Band limit (lower)
in_mode_01
none
Selected setpoint:
0 = SETP 1
1 = SETP 2
2 = SETP 3
3 = Last setpoint setting was carried out
through an external programmer
in_mode_02
none
Selftuning type:
0 = Selftuning „off“
1 = Selftuning „once“
2 = Selftuning „alwayS“
in_mode_03
none
Type of the external programmer input:
0 = Voltage
0 V to 10 V
1 = Current
0 mA to 20 mA
in_mode_04
none
Internal/external temperature control:
0 = Temperature control with internal sensor.
1 = Temperature control with external Pt100 sensor.
in_mode_05
none
Circulator in Stop/Start condition:
0 = Stop
1 = Start
in_mode_08
none
Adjusted control dynamics
0 = aperiodic
1 = standard
in_hil_00
none
Max. cooling power (%).
in_hil_01
none
Max. heating power (%).