Atc sensor - internal / external, Atc status - yes / no – JULABO SE-Z Heating Circulators User Manual
Page 51
HE / SE
51
Level 2
Parameter level
•
Integer digits flash, set by pressing
+
•
Decimal digits flash, set by pressing
+
If only a 2-point calibration is carried out, the following menu items are not
indicated anymore.
The value is only indicated
•
Integer digits flash, set by pressing
+
•
Decimal digits flash, set by pressing
+
9.7.1.
ATC SENSOR - INTERNAL / EXTERNAL
In the first submenu the ATC function is set for the >INTERN< internal or the
>EXTERN< external temperature sensor.
Calibration can be carried out for the internal temperature sensor and for the
external temperature sensor connected to the socket „ext. Pt100“.
The circulator is able to save both parameter sets. However only the one
which has been set under menu item >ATC SENSOR < is displayed.
9.7.2.
ATC STATUS - YES / NO
In the second submenu the ATC function for the temperature sensor selected
above is activated >YES< or deactivated >NO<.
>YES< (factory setting) The controller of the circulator uses the original
curve of the temperature sensor or the new curve measured during the
ATC calibration.
Important: Set to >NO< during the calibration process
>NO< An ATC calibration is to be carried out.
Important: Set to >YES< after calibration.
In the > ATC STATUS < >YES< the ATC calibration always affects the
current working temperature; also the one set via interface.