Menu config - configuration of unit, E 38 – JULABO SE-Z Heating Circulators User Manual
Page 38
Menu functions
38
9.4.
MENU CONFIG - Configuration of unit
Menu level 1
A RESET can be effected only in the >OFF< mode.
Switch off circulator by pressing the
key and call up the menu
CONFIGURATION.
Level 2
Parameter level
Press the key
if a parameter is to be
retained. Correction function for parameters and
values (prior to OK).
or
•
Switch on and off remote control by pressing
and
Control display in the topline
for Remote
For remote control refer to 71
Connect RS232 with PC.
or
or
•
Switch over setpoint setting by pressing
and
OFF –Setpoint setting with the navigation keys
or
Setpoint setting via the analog socket „ext. Pt100
or
analog socket >REG+EPROG<
or
•
Switch on and off autostart by pressing
and
AUTOSTART on = on
AUTOSTART off = off
See WARNING page 40
or
•
Switch on and off OFFMODE by pressing
and
PUMP ON
continuous operation of circulating pump
PUMP OFF circulating pump is linked to Start/Stop
or
or
•
Switch over the input variable by pressing
and
Programming of variables for the parameters >
SERIAL < or > EPROG < is only accepted, if the
unit is in Start mode