Menu pump - setting of pump pressure, E 37 – JULABO SE-Z Heating Circulators User Manual

Page 37

background image

HE / SE

37

9.3.

MENU PUMP - Setting of pump pressure

Examples:

Soll

Ist

Factory setting:

stage 1

The pressure of the circulating pump is adjustable in four stages. After
setting, the VFD COMFORT-DISPLAY indicates the corresponding value.

Adjustable pump capacity

stage 1 ... 4

Illuminated display:

for pump pressure

Adjusted:Display for the adjusted pump pressure stage in the –OFF-
mode.

Effective: Display for the effective pump pressure stage (rotation
speed) after start.

For protecting the pump motor, the rotation speed (i.e. the pump pressure
stage) changes with the load applied.
Example: The viscosity of the bath fluid (i.e. the load applied to the pump
motor) changes with the working temperature in the bath.

Flow rate:

22 ... 26 l/min

Pump capacity stage

1

2

3

4

Pump pressure

[bar]

0.4

0.5

0.6

0.7

Suction pump

[bar]

0.2

0.26

0.33

0.4

Total capacity

[bar]

0.6

0.76

0.93

1.1

in a loop circuit

1. Press the

key.

2. Select the menu >PUMP< pressing the

key and confirm by

pressing the

key

The set parameter flashes (example: >LEVEL 2<)

3. Change the parameter by pressing

and confirm by pressing the

key.

or

Press the

key if the parameter is to be retained.