12.3.

List of commands

out commands:

Setting temperature values or parameters.

Command

Parameter

Response of circulator

out_mode_01 0

Use working temperature > SETPNT 1<

out_mode_01 1

Use working temperature > SETPNT 2<

out_mode_01 2

Use working temperature > SETPNT 3<

out_mode_04

0

Temperature control of internal bath.

out_mode_04

1

External control with Pt100 sensor.

out_mode_05

0

Stop the unit = R –OFF-.

out_mode_05

1

Start the unit.

out_mode_08

0

Set the control dynamics - aperiodic

out_mode_08

1

Set the control dynamics - standard

out_sp_00

xxx.xx

Set working temperature. „SETPNT 1“

out_sp_01

xxx.xx

Set working temperature. „SETPNT 2“

out_sp_02

xxx.xx

Set working temperature. „SETPNT 3“

out_sp_03 xxx.xx Set high temperature warning limit „OVERTMP“

out_sp_04 xxx.xx Set low temperature warning limit „SUBTMP“

out_sp_07

x

Set the pump pressure stage. (1 ... 4)

out_par_04

x.x

CoSpeed 0 ... 5.0 Band limit during external control.
Setting the maximum difference between the temperatures
in the internal bath and external system.

out_par_06

xxx

Xp control parameter of the internal controller.

0.1 ... 99.9

out_par_07

xxx

Tn control parameter of the internal controller.

3 ... 9999

out_par_08

xxx

Tv control parameter of the internal controller.

0 ... 999

out_par_09

xxx

Xp control parameter of the cascade controller.

0.1 ... 99.9

out_par_10

xxx

Proportional portion of the cascade controller.

1 ... 99.9

out_par_11

xxx

Tn control parameter of the cascade controller.

3 ... 9999

out_par_12

xxx

Tv control parameter of the cascade controller.

0 ... 999

out_par_13

xxx

Maximum internal temperature of the cascade controller.

out_par_14

xxx

Minimum internal temperature of the cascade controller.

out_par_15 xxx

Band limit (upper)

0 ... 200 °C

out_par_16 xxx

Band limit (lower)

0 ... 200 °C