beautypg.com

JULABO FP89-ME Ultra-Low Refrigerated Circulators User Manual

Page 11

background image

ME

11


Safety installations according to IEC 61010-2-010:
Excess temperature protection

adjustable from 0 °C ... 230 °C

Low liquid level protection

float switch

Classification according to DIN 12876-1

class III




Supplementary safety installations
Early warning system for low level

float switch

High temperature warning function

optical + audible (in intervals)

Low temperature warning function

optical + audible (in intervals)

Supervision of working sensor

plausibility control

Reciprocal sensor monitoring between
working and safety sensors

difference >35 K

Alarm message

optical + audible (permanent)

Warning message

optical + audible (in intervals)




Environmental conditions according to IEC 61 010-1:

Use only indoor.

Altitude up to 2000 m - normal zero.

Ambient temperature: +5 ... +40 °C

Air

humidity:

Max. rel. humidity 80 % for temperatures up to +31 °C,

linear decrease down to 50 % relative humidity at a temperature of +40 °C

Max. mains fluctuations of ±10 % are permissible.


Protection class according to IEC 60 529

IP21

The unit corresponds to Class I

Overvoltage category

II

Pollution degree

2

Caution:

The unit is not for use in explosive environment


Standards for interference resistance according to EN 61326-1
This unit is an ISM device classified in Group 1 (using high frequency for internal purposes) Class A
(industrial and commercial range).