JULABO FP89-ME Ultra-Low Refrigerated Circulators User Manual
Page 11
ME
11
Safety installations according to IEC 61010-2-010:
Excess temperature protection
adjustable from 0 °C ... 230 °C
Low liquid level protection
float switch
Classification according to DIN 12876-1
class III
Supplementary safety installations
Early warning system for low level
float switch
High temperature warning function
optical + audible (in intervals)
Low temperature warning function
optical + audible (in intervals)
Supervision of working sensor
plausibility control
Reciprocal sensor monitoring between
working and safety sensors
difference >35 K
Alarm message
optical + audible (permanent)
Warning message
optical + audible (in intervals)
Environmental conditions according to IEC 61 010-1:
Use only indoor.
Altitude up to 2000 m - normal zero.
Ambient temperature: +5 ... +40 °C
Air
humidity:
Max. rel. humidity 80 % for temperatures up to +31 °C,
linear decrease down to 50 % relative humidity at a temperature of +40 °C
Max. mains fluctuations of ±10 % are permissible.
Protection class according to IEC 60 529
IP21
The unit corresponds to Class I
Overvoltage category
II
Pollution degree
2
Caution:
The unit is not for use in explosive environment
Standards for interference resistance according to EN 61326-1
This unit is an ISM device classified in Group 1 (using high frequency for internal purposes) Class A
(industrial and commercial range).