Page 40
Menu functions
40
Level 2
Parameter level
0 … 999
The parameter flashes, set by pressing
+
0.1 … 99.9
The parameter flashes, set by pressing
+
9.5.1.
CONTROL – Control INTERNAL / EXTERNAL
Pt100
Switchover can only be effected if a Pt100 external sensor is
connected.
Factory setting: INT
IMPORTANT: Additional measures for external temperature control
Suggested settings for external temperature control:
BAND HIGH / LOW and INTERN MAX / MIN
see chapter >LIMITS< page 50.
Sensor calibration of the Pt100 external sensor is carried out in the
menu >ADJUST<, submenu >ATC SENOR - EXT<; set ATC
STATUS< to >OFF<
(See page 44).
Attention:
Place the external sensor into the temperature-controlled medium and securely fix the
sensor.
Pt100
M+R
Accessory: Pt100 external sensor
Order No.
Description
Material
Cable
8981003
200x6 mm Ø,
stainless steel
1.5 m
8981005
200x6 mm Ø,
glass
1.5 m
8981006
20x2 mm Ø,
stainless steel
1.5 m
8981010
300x6 mm Ø,
stainless steel
1.5 m
8981015
300x6 mm Ø,
stainless steel / PTFE coated
3 m
8981013
600x6 mm Ø,
stainless steel / PTFE coated
3 m
8981016
900x6 mm Ø,
stainless steel / PTFE coated
3 m
8981014
1200x6 mm Ø,
stainless steel / PTFE coated
3 m
8981103 Extension cable for Pt100 sensor
3.5 m
8981020 M+R in-line Pt100 sensor
The M+R in-line Pt100 sensor is a flow sensor and can be installed loop
circuit