Safety recommendations

 Employ suitable connecting tubing.

 Make sure that the tubing is securely attached.

 Avoid sharp bends in the tubing, and maintain a sufficient distance from

surrounding walls.

 Regularly check the tubing for material defects (e.g. for cracks).

 Preventive maintenance: Replace the tubing from time to time.


5.4. Filling/draining

Notice

 Pay attention to the thermal expansion of bath oil during heating to avoid

overflowing of the liquid.

 Do not drain the bath fluid while it is hot or cold!

Check the temperature of the bath fluid prior to draining (by switching the unit on
for a short moment, for example).

 Store and dispose of the used bath fluid according to environmental protection

laws.




Filling

Make sure that no bath fluid enters the interior of the circulator during
filling.

Recommended maximum filling level with water as bath fluid:
30 mm below the tank rim

Recommended maximum filling level with bath oils:
40 mm below the tank rim

After filling, immerse the samples in the bath, or place the lid on
the bath if the opening is not to be used.

The ME circulator has an early warning system for low level that
may be triggered when changing samples in the bath.

Draining:

 The drain tap is located behind the venting grid on the front of

the unit, which can be taken off by removing 4 screws.

 Slide a short piece of tube onto the drain connection and hold it

into a container.

 Loosen the drain screw a few turns and drain the unit

completely.

 Tighten the drain tap.