JULABO FP89-ME Ultra-Low Refrigerated Circulators User Manual
Page 55

ME
55
Short circuit of control line to cooling machine during self-test.
Error message with ticker:
> SAFETY SENSOR ALARM-CALL SERVICE <
The cable of the excess temperature protector has been disconnected or
short-circuited
Error message with ticker:
> LOW LEVEL WARNING-FILL MEDIUM <
The early warning system for low level reports a critical fluid level. Refill bath
fluid.
60
30
15
45
4 s
By quickly switching off and restarting the unit the alarm is cancelled.
If the error occurs once more after the restart, a remote diagnosis is required.
Error message with ticker:
> CONFIGURATION ERROR-PRESS OK<
The configuration of the circulator does not correspond with its current
application.
Press the
key for a non-recurring, automatic change of the
configuration.
In this case please call the JULABO Technical Service or an authorized dealer.
Disturbances that are not indicated.
The electronic pump motor is overload-protected by an electronic current
limiter. If viscosity of the bath fluid is or becomes too high, the motor stops
running.
15
A
M
P
Mains circuit breakers (resettable) 15 A
Warning:
Always turn off the unit and disconnect the mains cable from the power source
before performing any service or maintenance procedures.
Electrical connections and any other work must be performed by qualified
personnel only.