JULABO FP89-ME Ultra-Low Refrigerated Circulators User Manual

Page 55

background image

ME

55

Short circuit of control line to cooling machine during self-test.

Error message with ticker:
> SAFETY SENSOR ALARM-CALL SERVICE <

The cable of the excess temperature protector has been disconnected or
short-circuited

Error message with ticker:
> LOW LEVEL WARNING-FILL MEDIUM <

The early warning system for low level reports a critical fluid level. Refill bath
fluid.

60

30

15

45

4 s

By quickly switching off and restarting the unit the alarm is cancelled.
If the error occurs once more after the restart, a remote diagnosis is required.

Error message with ticker:
> CONFIGURATION ERROR-PRESS OK<

The configuration of the circulator does not correspond with its current
application.

Press the

key for a non-recurring, automatic change of the

configuration.
In this case please call the JULABO Technical Service or an authorized dealer.


Disturbances that are not indicated.

The electronic pump motor is overload-protected by an electronic current
limiter. If viscosity of the bath fluid is or becomes too high, the motor stops
running.

15

A
M
P

Mains circuit breakers (resettable) 15 A

Warning:

 Always turn off the unit and disconnect the mains cable from the power source

before performing any service or maintenance procedures.

 Electrical connections and any other work must be performed by qualified

personnel only.