beautypg.com

Start-menu – JULABO FP89-ME Ultra-Low Refrigerated Circulators User Manual

Page 31

background image

ME

31

9.1.

MENU PROGRAM – START

Start-Menu

This menu will start a previously set program.

Requirements:

1. Create a program. (refer to next chapter)

2. Return to the Start-MENU and confirm the desired setting of each

MENU item with the key

3. Set a start time (>TIME< >DATE< >YEAR<) if the program is to be

started by the internal timer.

Menu level 1

> STEP<

Program start at section 1 … 10

> RUNS <

Number of repetitions 1 ... 99

> END<

Status at end of program (STDBY/SETPNT)

Standby or last setpoint

> GO <

Time of start (NOW/TIMER)

Level 2

Parameter level

Press the

key if a parameter is to be retained.

Correction function for parameters or values (prior to
OK)

Set program step with

and

example: STEP 1

Set number of runs with

and

example: 1 run

or

Set desired parameters with

and

.

(STDBY / SETPNT)
Parameter STanDBY: the circulator switches to –
OFF-.
Parameter SETPoiNT: the circulator constantly
keeps the temperature at the value of the last step.

or

Confirm >NOW< with the

key and the program

will start immediately

or
start at the set time under parameter (TIMER ).
Set time in the example below:
09. August 2009, 11:15 hrs