JULABO FCW1600_1600S Recirculating coolers User Manual
Page 36

Menu functions
36
10.3. Adjustable
parameters
ACTUAL
SET
Set the parameters for the following menu items in the
configuration level:
OP - Operating mode
0
=
Keypad
control
1
= Remote control via RS 232
ACTUAL
SET
HAnd - Handshake of the serial interface
0
= XOn/XOff, software handshake
1
= RTS/CTS, hardware handshake *
ACTUAL
SET
PAr - Parity bits of the serial interface
0
=
no
1
=
odd
2
=
even
*
ACTUAL
SET
bAud - Baud rate of the serial interface
1200
Bauds
2400
Bauds
4800 Bauds*
9600
Bauds
ACTUAL
SET
E_Sb - External stand-by for emergency switch-off
0
= stand-by input is ignored *
1
= stand-by input is active
(Stand-by input - see page 30)
(* factory setting)