Recirculating coolers
21
5.7. Draining
Notice:
Do not drain the bath fluid while it is hot or cold!
Check the temperature of the bath fluid prior to draining (by switching the
unit on for a short moment, for example).
Store and dispose of the used bath fluid according to the environmental
protection laws.
MIN
MAX
0
1
2
3
4
5
6
1.
MIN
MAX
0
1
2
3
4
5
6
1.
9.
8.
Turn the mains switch (1.) off.
Hold the venting grid, pull out and remove.
Take the drain tubing (8.) out of the holder and hold it into
a pail.
Open the drain tap (9.) and empty the unit completely.
Close the drain tap and replace the drain tubing into the
holder.
Replace the venting grid.
MIN
MAX
0
1
2
3
4
5
6
9.
8.
1.
In case the recirculating cooler is placed on the floor, the
unit may be drained using a suction pump unit.
Connect the drain tubing (8.) to the suction pump unit.
Open the drain tap (9.).
Switch the pump on and fully empty the unit.