Operator responsibility – Safety instructions
10
Safety installations
(adjustable via LED):
High temperature protection
°C -25 ... 85
Low temperature protection
°C -25 ... 85
Low liquid level protection
float switch
Classification according to DIN 12876-1
III
Overload protection for pump motor
contactor
Overload protection for cooling compressorcontactor
Alarm signal
optical + audible
Environmental conditions according to IEC 61 010-1:
Use only indoor.
Altitude up to 2000 m - normal zero.
Ambient temperature: +5 ... +40 °C
Air
humidity:
Max. rel. humidity 80 % for temperatures up to +31 °C,
linear decrease down to 50 % relative humidity at a temperature of +40 °C
Max. mains fluctuations of ±10 % are permissible.
The unit corresponds to Class I
Overvoltage category
II
Pollution degree
2
Caution:
The unit is not for use in explosive environment.
Standards for interference resistance according to EN 61326-1
This unit is an ISM device classified in Group 1 (using high frequency for internal purposes)
Class A (industrial and commercial range).