Recirculating coolers

23

6.2.1. Automatic / non-automatic start mode

NOTE:
The recirculating cooler has been configured and supplied by JULABO according to
N.A.M.U.R. recommendations. This means for the start mode, that the unit must enter a safe
operating state after a power failure (non-automatic start mode). This safe operating state is
indicated by „OFF“ or „rOFF“, resp. on the MULTI-DISPLAY (LED). A complete shutdown of
the main functional elements such as heater and circulation pump is effected simultaneously.
Should such a safety standard not be required, the AUTOSTART function (automatic start
mode) may be activated, thus allowing the start of the recirculating cooler directly by pressing
the mains power switch or using a timer.

Activating/deactivating AUTOSTART

1. Turn on the recirculating cooler with the mains power

switch
and
press the Enter key to start operation.

2. Simultaneously press the safety temperature key (C)

and the MENUE key (4.2) to enter the setting mode.

Press the edit key

to select the parameter on the

MULTI-DISPLAY (LED2).
„1“ - AUTOSTART off.
„0“ - AUTOSTART on.

Press the Enter key

to store the parameter.

3. Simultaneously press the safety temperature key (C)

and the MENUE key (4.2) to exit the setting mode.

The AUTOSTART function (automatic start mode)
allows the start of the recirculating cooler directly by
pressing the mains power switch or using a timer.

Warning:
For supervised or unsupervised operation with the AUTOSTART function,
avoid any hazardous situation to persons or property.
The recirculating cooler does no longer conform to N.A.M.U.R.
recommendations.

Take care you fully observe the safety and warning functions of the