beautypg.com

JULABO FPW50-ME Refrigerated and Heating Circulators User Manual

Page 9

background image

ME

9

2.4. Technical

specifications

F25-ME

Working temperature range

°C

-28 ... 200

Temperature stability

°C

±0,01

Cooling capacity

Medium ethanol

°C
kW

+20 0 -20
0.26 0.2 0.06

Refrigerant

R134a

Overall dimensions

(WxDxH)

cm

23x42x61

Bath opening

(WxL)

cm

12x14

Bath depth

cm

14

Filling volume

from ... to

liters

3 ... 4,5

Weight

kg

31

Mains power connection

230 V/50 Hz

V/ Hz

207-253 / 50

Current draw

(at 230 V)

A

12

Current draw

(at 230 V) CH

A

<9 + 2>

Mains power connection

230 V/60 Hz

V/ Hz

207-253 / 60

Current draw

(at 230 V)

A

12

Mains power connection

115 V/60 Hz

V/ Hz

103-127 / 60

Current draw

(at 115 V)

A

13

Mains power connection

100 V/50-60 Hz

V/ Hz

90-115 / 50-60

Current draw

(at 100 V)

A

13


F26-ME

F32-ME

Working temperature range

°C

-28 ... 200

-30 ... 200

Temperature stability

°C

±0,01

±0,01

Cooling capacity

Medium ethanol

°C
kW

+20 0 -20
0.26 0.2 0.06

+20 0 -20 -30

0.45 0.39 0.15 0.05

Refrigerant

R134a

R134a

Overall dimensions

(WxDxH) cm 42x42x42

31x42x64

Bath opening

(WxL)

cm

12x14

18x12

Bath depth

cm

14

15

Filling volume

from ... to

liters

3 ... 4,5

5,5 ... 8

Weight

kg

31

37

Mains power connection

230 V/50 Hz

V/ Hz

207-253 / 50

207-253 / 50

Current draw

(at 230 V)

A

12

12

Mains power connection

230 V/60 Hz

V/ Hz

------

207-253 / 60

Current draw

(at 230 V)

A

------

12

Mains power connection

115 V/60 Hz

V/ Hz

103-127 / 60

103-127 / 60

Current draw

(at 115 V)

A

13

14

Mains power connection

100 V/50-60 Hz

V/ Hz

-----

90-110 / 50-60

Current draw

(at 100 V)

A

-----

14

All measurements have been carried out at:
rated voltage and frequency

ambient temperature: 20 °C

Technical changes without prior notification reserved.