beautypg.com

JULABO FPW50-ME Refrigerated and Heating Circulators User Manual

Page 34

background image

Menu functions

34

Submenu TIMER

Parameter level

set the time for the start of the program in the
submenu >TIMER<.

>TIME< hours/minutes (hh:mm), set both values one
after the other and confirm

hours flash, set by pressing

+

minutes flash, set by pressing

+

>DATE< day/months (TT/MM), set both values one after
the other and confirm.

day flashes, set by pressing

+

month flashes, set by pressing

+

>YEAR< year

Set the year with

and

.

The program starts at the set time.

Display of time until start:

In line 3 the notice >TIMER< and the set values for „TIME“ and
„DATE/YEAR“ are alternately indicated

Check the correct setting of the internal real time clock if required
(see MENU CONFIG)


The started program


A

B

C1

After the start the program will indicate the currently calculated setpoint in line
2
S XX.XX. The value increases within the time period >TSLICE< until the
target temperature >SETPNT< of the section is reached.
If the time period in a section is set to „0“, the next section will not begin until
the target temperature has been reached.

Use the edit keys

to scroll to line 3. The display changes

approximately every 4 seconds between the current section (STEP XX) and
the

A remaining time of the section
B remaining time of the program
C current bath temperature

I xxx.xx - internal actual value or