JULABO FPW50-ME Refrigerated and Heating Circulators User Manual
Page 35
ME
35
C2
D1
D2
E xxx.xx – external actual value
D RUN – the program has started or
PAUSE – the progress of the program has been interrupted by pressing
the
key. While the time is stopped the temperature will constantly
remain at the last calculated setpoint
Continue with the
key.
Termination / Interruption of a program
The program can be terminated any time by pressing the key
.
In case of power failure the program is interrupted.
The circulator switches to –OFF- .
If the AUTOSTART-function is activated the programmer starts again at
the point in time approx. 5 minutes prior to the interruption. However, an
uncontrolled change of the bath temperature has occurred.
9.2.
MENU PROGRAM – creation, administration
Menu level 1
1 program
10 sections
The integrated programmer permits fast and easy programming of setpoint
temperature sequences. This temperature sequence is called program. A
program is composed of individual sections (STEP). The sections are defined
by duration (TSLICE) and target temperature. The target temperature is the
setpoint (SETPNT), which is achieved at the end of a section. The programmer
calculates the temperature ramp from the difference in time and temperature.
STEP
(Nr.)
1 2 4 6 8
SETPNT
(°C) 100 180 180 75 75
TIME (hh.mm)
00:20 00.10 00:20 00:50 00:20
Graph1
25
50
10 20 30 40 50 60 10 20 30 40 50 60 10 [min]
75
100
150
200
°C
t
Step 1
2
4
6
8
Sections without set value and time are skipped. They can be defined
retroactively and the integrated into the program.
- FP50-ME Refrigerated and Heating Circulators FP40-ME Refrigerated and Heating Circulators F34-ME Refrigerated and Heating Circulators F33-ME Refrigerated and Heating Circulators F32-ME Refrigerated and Heating Circulators F26-ME Refrigerated and Heating Circulators F25-ME Refrigerated and Heating Circulators