Cleaning / repairing the unit – JULABO FPW50-ME Refrigerated and Heating Circulators User Manual
Page 66

Cleaning / repairing the unit
66
14. Cleaning / repairing the unit
Caution:
Always turn off the unit and disconnect the mains cable from the power source
before cleaning the unit.
Prevent humidity from entering into the circulator.
Electrical connections and any other work must be performed by qualified
personnel only.
Maintaining the cooling performance
To maintain the full cooling performance, clean the condenser from time to
time.
Switch off the unit, disconnect mains power cable.
Hold the venting grid, pull out and remove.
Clean the ribbed condenser with a vacuum cleaner.
Replace the venting grid.
Switch on the unit.
Cleaning:
For cleaning the bath tank and the immersed parts of the circulator, use low
surface tension water (e.g., soap suds).
Clean the outside of the unit using a wet cloth and low surface tension water.
The circulator is designed for continuous operation under normal conditions.
Periodic maintenance is not required.
The tank should be filled only with a bath fluid recommended by JULABO. To
avoid contamination, it is essential to change the bath fluid from time to time.
Repairs
Before asking for a service technician or returning a JULABO instrument
for repair, please contact an authorized JULABO service station.
When returning the unit:
Clean the unit in order to avoid any harm to the service personnel.
Attach a short fault description.
During transport the unit has to stand upright. Mark the packing
correspondingly.
When returning a unit, take care of careful and adequate packing.
JULABO is not responsible for damages that might occur from insufficient
packing.
JULABO reserves the right to carry out technical modifications with repairs for
providing improved performance of a unit.
- FP50-ME Refrigerated and Heating Circulators FP40-ME Refrigerated and Heating Circulators F34-ME Refrigerated and Heating Circulators F33-ME Refrigerated and Heating Circulators F32-ME Refrigerated and Heating Circulators F26-ME Refrigerated and Heating Circulators F25-ME Refrigerated and Heating Circulators