Supply

DANGER

The gauge may only be connected to power sup-
plies, instruments or control devices that conform
to the requirements of a grounded protective extra-
low voltage (SELV) and limited power source
(LPS), Class 2. The connection to the gauge has to
be fused

3)

.

Supply voltage

at the gauge

Class 2 / LPS
+14 … +30 VDC or
±15 V (±5%)

Ripple

≤1 V

pp

Power consumption
while being heated
at

operating

temperature


≤12 W
≤8 W

The gauge is protected against reverse polarity of the supply
voltage and overload.

Electrical connection

3CC1-xxx-0xxx
3CC1-xxx-2xxx


9-pin D-Sub, male
15-pin D-Sub, male

Sensor cable for

3CC1-xxx-0xxx
3CC1-xxx-2xxx


9-pin plus shielding
15-pin plus shielding

Sensor cable

15 poles plus shielding

Cable length

Supply voltage 15 V


≤ 8 m (0.14 mm²/conductor)
≤15 m (0.25 mm²/conductor)

Supply voltage 24 V

≤43 m (0.14 mm²/conductor)
≤75 m (0.25 mm²/conductor)

Supply voltage 30 V

≤88 m (0.14 mm²/conductor)
≤135 m (0.25 mm²/conductor)

3)

INFICON controllers fulfill this requirement.