INFICON SKY CDG045D 0.05 ... 1000Torr / mbar User Manual

Page 10

background image

10

tina51e1-c (2011-12)

Remote Zero Adjust


External switching contact

digital input for zero adjust-
ment (

→  23)

30 VDC / <5 mA DC


Switching functions

SP1, SP2

Setting range

0 … 99% F.S. (0 … 9.9 V)

Hysteresis 1%

F.S.

Relay contact

30 VDC/ ≤0.5 ADC
floating (n.o.)

closed p

≤ p

SP

(LED lit solid)

open p

≥ p

SP

(LED off)

Switching time

≤50 ms


Status relay

Relay contact

30 VDC/ ≤0.5 ADC
connected to supply com-
mon (pin 5)

closed measurement

mode

warning

open no

supply voltage

warming up
error


RS232C interface

Transmission rate
Data format


9600 baud
binary
8 data bits
one stop bit
no parity bit
no handshake
→ "Electrical Connection"

For further information on the RS232C interface

→  [4].


Diagnostic port

Jack connector, 2.5 mm,
3-pin