4 crystal replacement instructions – INFICON Cool Drawer Single Sensor User Manual
Page 50

4 - 10
PN
07
4-
60
9-
P1
A
Cool Drawer Single and Dual Sensors Operating Manual
4.4 Crystal Replacement Instructions
CAUTION
To preserve cleanliness and to maximize crystal
performance, all work should be performed in a clean
room environment.
1
Using a thumb and index finger, gently squeeze the sides of the retainer at
mid-section, then lift the retainer up, away from the drawer, as shown
in
2
Hold the drawer by the handle and turn it upside down to remove the spent
crystal.
3
Prior to installing the new crystal, review
section 4.3.1, Handle the Crystal with
4
Wear clean nylon gloves and grasp the edge of the new crystal with clean
plastic tweezers.
5
Orient the new crystal so the patterned electrode is facing up.
6
Insert the new crystal in the drawer and release the crystal. The pattern
electrode must be facing up as shown in
7
Hold the retainer by its sides. Align its orientation notch with the drawer then
gently and evenly push the retainer down until it snaps firmly into the drawer.
See
CAUTION
Never push down (or pull up) on the contact spring, doing
so may permanently damage it.
8
Inspect the entire assembly. The retainer should lay evenly and engage the
drawer at all four corners.
9
Reinstall the drawer into the sensor body. Push the drawer straight in making
certain that it is completely seated in the sensor body.