INFICON UL1000 Helium leak detector Interface description User Manual
Page 9

Interface Description
1-5
iin
s
70
e1
.fm
Op
er
at
in
g M
a
nua
l
(1
00
4)
*MEAS
:P1
inlet pressure
:P2
foreline pressure
:P3
inlet pressure (UL5000)
:OFFset
offset current [A]
:IMess
current raw values [A]
:IFilter
current filtered [A]
:UNV
amplifier voltage [V]
:UVV
preamplifier voltage [V]
:MIAP
anode potential [V]
:MIKP
cathode potential [V]
:MISP
suppressor potential [V]
:MIAKP
anod-/cathode potential [V]
:VALVE
valve voltage [V]
:U24EXT
external voltage [V]
:U24FB
voltage remote contol [V]
:U24FAN
fan voltage [V]
:TEMPeratur
:Amplifier
preamplifier temperature [°C]
:Electronic
electronic temperature [°C]
:TURBO
:Frequency
TMP frequency [Hz]
:Voltage
TMP voltage [Hz]
:Current
TMP current [A]
:Power
TMP power [W]
:BOOSTER
Booster-TMP frequency (UL5000 only)
*CONFig
:ALARMDelay
alarm delay after evacuation (0 ... 600, 65535)
:AUDio
audio alarm type (PIN, SET, TRIG, PROP)
:BACKGround
background display (ON, OFF)
:BEEP
beep-sound (ON, OFF)
:BOOster
Booster-TMP Mode: ON, OFF
:CALAccess
CAL access (ON, OFF)
:CALleak
leak rate of test leak
:INT
internal test leak
:EXTVAC
external test leak in vacuum mode
:EXTSNIFF
external test leak in sniff mode
:CALREQ
calibration request
:CAThode
cathode
:CONTrol
control location (LOCAL, RS232, PLC, LOCAL/RS232,
LOCAL/PLC)
:FREQuenz
mains frequency (50/60Hz)
:HYDros
HYDROS: DISABLED, ON, OFF
:ICAL
filter type
:LANGuage
language (ENGlish, DEUtsch, FRAncais, ITAliano,
POLski, KATakana, CHInese, ESPanol, KORean)
:LCDAutorange
display range (auto, manual)
:LCDContrast
display contrast (0 ... 99)
:LCDDECades
number of display decades (2 ... 9)
:LCDInvert
invert display
:LCDSCALELog
display scale
:LIMITLow
lower display limit (0 ... 7)
:MASS
mass (2, 3, 4)
:MFAE:
anode potential reference