1.7
Command list
*CLS
clear Error
*IDN
identification
:CRC
check sum
:DEVice
name of instrument (UL1000, UL1000Fab, UL5000)
:VERsion
software version
:SERial
serial-Number
:TURBO
software version TC600
:MC68
hardware identification MC68
:IOversion
hardware identification IO-Karte
:GBversion
hardware identification control panel
:VDversion
hardware identification mother board
:DIP1
MC68 DipSwitch 1
:DIP2
MC68 DipSwitch 2
:TURBO
software version frequency converter
*STATus
mode of calibration routine
INIT, ACCL, STBY, VENT, WAIT_EVAC, EVAC,
MEAS, CAL, ERROR
[:CAL]
mode calibration
IDLE, EVAC, OPEN, TUNE, TUNE_RES, CLOSE,
STABLE_CLOSE, WAIT_OK
[:CALHist]
CAL History (1…12)
[:CALMode]
kind of calibration INT_AUTO, INT_MAN, EXT
[:ERRor]
error
[:ERRorHist]
error History (1…12)
[:ZERO]
zero
[:RANGE]
measuring mode GROSS, FINE, ULTRA
:BOOSTER
Booster-TMP State (UL5000 only): OFF, ACCL,
NORM, FAIL
:RESULT
AutoLeakTest state/result: IDLE, RUNNING, PASS,
FAIL
:TESTLog
Test-Log
*READ
leak rate (selected unit)
[:
leak rate (desired unit)
*STArt
start
*STOp
stop
*PURGE
Purge on
:OFF
Purge off
*VENt
vent
*CAL
start calibration / proceed (see Chapter 1.8)
*ZERO
zero
[:OFF]
zero off