6.1.5 Administer settings
(save, load, ...)

All of the measurement channel settings
made by you in the HMG for the last meas-
urement can be saved and then reloaded
later in order to be able to repeat the same
measurement periodically, for example.

You can also delete individual stored set-
tings or restore the factory settings.

Select Administer settings (save, load, ...),
press OK, and now proceed to the Adminis-
ter Settings
menu with its four submenus
(see screenshot).






When you confirm Save current settings
with OK, you see the name of the setting
saved last, or No Name if nothing was
saved. Confirm with OK, and you are shown
a list of all the saved settings and/or the op-
tion of entering a new name. Select New
Name
and enter a name using the combina-
tion keypad.
Press OK x 2 to save the new name and to
return to the Administer Settings menu.
The HMG automatically assigns a consecu-
tive number, which can also be changed by
you.
The HMG automatically assigns the next
highest unused consecutive number, which
is always unique. Any gaps in the list of
numbers are not filled.