HYDAC ETS 300 User Manual
Page 10

9. Technical specifications
Input data for built-in sensor:
Measuring range:
-25 .. 100 °C (-13 .. 212 °F)
Pressure resistance:
600 bar
Overload pressure:
900 bar
Hydraulic connection:
G1/2 A DIN 3852
Torque rating:
45 Nm
Parts in contact with medium:
stainless steel, seal: FPM
Input data for separate sensor:
Display range: *)
-30 .. 150 °C (-22 .. 302 °F)
Sensor connection:
connector M 12x1, 4-pole
Output data:
Accuracy (display, analogue output) £ ±1.0 °C (£ ±2.0 °F)
Temperature drift:
£ ±0.15 % /10 K (zero point and range)
Reaction time (T09):
approx. 20 s
Analogue output:
4 .. 20 mA, ohmic resistance
£ 400 W
Switching outputs:
Type:
PNP transistor output
Switching current:
max. 1.2 A
Switching cycles:
> 100 million
Ambient conditions:
Temperature range of medium:
-25 .. + 100 °C (built-in sensor)
Ambient temperature range:
-25 .. + 80 °C
Storage temperature range:
-40 .. + 80 °C
Nominal temperature range:
-10 .. + 70 °C
- mark:
EN 50081-1, EN 50081-2
EN 50082-1, EN 61000-6-2
Vibration resistance:
approx. 10 g / 0..500 Hz
Shock resistance:
approx. 50 g / 1ms
Other data:
Supply voltage:
20 .. 32 VDC
Current consumption:
approx. 100 mA (no switching output)
Safety type:
IP 65
Material of housing:
tube: stainless steel
keypad housing: PA6.6 Gf30
Display:
3-digit, 7-segment LED, red,
height of digits 9.2 mm
Weight:
approx. 300 g
Technical specifications - Accessories
Temperature sensor TFP 100
Temperature range of medium (for TFP 100):
-40 .. +125 °C / -40 .. +257°F
Electrical connection:
4-pole Binder plug, series 714 M18
Safety sleeve for tank-mounting the TFP 100
Parts in contact with medium:
All materials compatible with nickel
Note: *) Depending on the temperature range of the temperature sensor fitted, this can
reduce the display range of the ETS 300 with a separate sensor input.