beautypg.com

Tests – Despatch RAD1-42-2E with Protocol Plus User Manual

Page 60

background image

52

Tests


Tests should be performed carefully and regularly. The safety of personnel as well as
the condition of equipment may depend upon the proper operation of any one of the
functions of the temperature control. Test the control every 40 hours. Check that the
heater LED is cycling on and off, indicating that the heater is working. Also check the
high limit function to make sure it is working properly.

To test the high limit:

1.

Go to Manual Mode and enter control setpoint value at least 20 F (11 C) lower
than the current process temperature.


2.

Press the Menu key and lower the high limit setpoint to a value just below the
current process temperature.


3.

The high limit alarm indicator should flash and a high limit alarm message should
be displayed. Verify that the heater relay has been disabled by checking that
2LED on the control panel circuit board is not lit.


4.

Return the control setpoint and high limit setpoint values to their original values.


5.

Press the Reset key.