19
Measurement Modes
The PX-7 is equipped with three measurement mode options (echo -
echo, interface-echo, or auto). In echo-echo mode, the gauge has the
ability to read thin metals down to .006 inches (.15 millimeters). The echo -
echo mode also allows the user to measure the thickness of metals that
have been previously coated or painted on the surface. This enables the
user to determine the thickness of the metal without having to remove the
paint. In interface-echo mode, the gauge has the ability to read plastics
and thicker materials. The PX-7's auto mode will automatically switch
between modes according to the different materials being measured.
Finally, the plas mod can be used to measure thin plastics using a special
graphite tip. The following section outlines how to switch between
measurement modes:
Measurement Modes
1) Press ON/OFF key to power up the PX-7 DL.
2) Press the MODE key to activate features and settings.
3) Press the UP or DOWN arrow keys to scroll to GAtE.
4) Press the SEND key to toggle between the different measurement
modes ( E-E, I-E, AutO, and PLAS ).
5) Press the MODE key once again to return to measurement mode.