Caron 6017 User Manual
Page 30
6010 Series Operations Manual
Rev E 2/14/2012
Page 30 of 68
Operation of the Defrost System (DEFR301)
A Defrost System is used to maintain operational temperature below 5ºC. When the
chamber temperature is below 4.5ºC, the defrost system is enabled automatically.
When the chamber temperature is above 4.5ºC (except during a defrost cycle), the
defrost system is disabled automatically.
When the defrost system is enabled (below 4.5ºC), a defrost cycle will occur once every
12 hours and lasts for 20 minutes. During the defrost cycle, the chamber temperature
will increase several degrees above set point to melt any ice that has formed on the
evaporator. This temperature ‘spike’ is normal and assures long-term operation.