Stacking two Units
The CARON 6010 family is designed to allow two units to be stacked. When units are
stacked they must be bolted together for safety. A stacking adaptor kit, STCK301 with
instructions is available through CARON customer service.
Leveling the Unit
Place a level on the middle shelf of the environmental chamber. Adjust the cabinet
leveling feet so the shelf is level. Units equipped with optional casters (CSTR101) can
be leveled by adjusting the height of the lock nut on the caster. Adjust the feet or
casters appropriately until the unit sits level left to right and front to back.
Connecting the Drain Line
The chamber drain connection is located in the bottom middle of the back of the
chamber. A 3/8” NPT fitting and tubing are supplied in the unit parts kit. Thread the
fitting into the drain connection and slide the tubing into the drain connection. Pull on
the tubing after installation to make sure it is tight. Route the drain tubing to a local floor
drain.
When using a pressurized water source, failure
to connect the unit to a drain could result in
facility flooding.
Failure to install the stacking adaptor kit can
result in the top unit falling causing serious
injury or death.
Do not have multiple loaded shelves out
simultaneously or the chamber may tip.