Caron 6017 User Manual
Page 27
6010 Series Operations Manual
Rev E 2/14/2012
Page 27 of 68
SECTION 5 – OPERATION
With the chamber properly installed and the appropriate utilities connected, the power
switch on the upper left of the control bezel can be turned on. Within a few minutes, the
temperature and humidity will begin to approach set-points. Allow the unit to stabilize for
12 hours before use or prior to making any calibration adjustments.
Changing the Temperature Set-point
The standard cabinet has a temperature set-point range of 5C to 70C. With the
purchase of and optional defrost package (DEFR201), the low end temperature range
can be extended to 0C.
To set the temperature set-point, press the UP arrow to increase the temperature set-
point by 0.1
C. Press the DOWN arrow to decrease the temperature set-point by 0.1
C.
Pressing and holding either button will cause the set-point to scroll rapidly in either
direction.
To turn off the temperature control system, but still display the chamber temperature
level, press the EZ button. The words “OFF” will display in the set-point area. To toggle
back to controlled temperature, press the EZ button again.
If the unit is equipped with the deluxe controller package (DLUX301 or DLUX302),
additional ramp and soak features are available. Refer to Section 7 of the manual to
program these features.
Temperature
Set-point
Actual
Temperature
Increase
Temperature
Set-point
Decrease
Temperature
Set-point
EZ Button