Caron 6016 User Manual
Page 47
6014 & 6016 Operations Manual
Rev C 2/14/12
Page 47 of 56
SECTION 9
– PREVENTATIVE MAINTENANCE
The CARON incubator has been robustly designed to minimize performance problems.
However, regular maintenance is very important for continuous trouble free operation.
As a general rule, CARON recommends an annual calibration check of the temperature,
humidity and CO2 systems. CARON offers a full range of on-site calibration and
validation services. We also offer preventative maintenance contracts on our
equipment. Contact our service department for at 740-373-6809 details or visit us on
the web at
Recommended Daily Maintenance Checks
Check the Temperature, humidity, and CO2 displays versus set-points.
Check for and correct any alarm condition.
Recommended Monthly Maintenance Checks
Check to ensure the drain in the bottom of the unit is draining properly.
Check front air intake filter. If it is dirty replace it with CARON part number FLTR303.
Washing the filter will result in poor performance.
Recommended Annual Maintenance Checks
Disinfect all interior surfaces with a general purpose laboratory cleaning agent.
Perform a complete calibration of the temperature and humidity systems.
A full validation is recommended for GMP facilities each time a unit is installed,
moved or
undergoes significant repair. Contact CARON’s service department to
schedule on-site validation.