Caron 6016 User Manual
Page 24

6014 & 6016 Operations Manual
Rev C 2/14/12
Page 24 of 56
SECTION 5
– OPERATION
With the incubator properly installed and the appropriate utilities connected, the power
switch on the upper left of the control bezel can be turned on. Within a few minutes, the
temperature and humidity will begin to approach set-points. Allow the unit to stabilize for
12 hours before use or prior to making any calibration adjustments.
Changing the Temperature Set-point
If an alternative temperature set-point is required, the following steps can be taken:
Press the up arrow to increase the temperature set-point by 0.1
C. Press the down
arrow to decrease the temperature set-point by 0.1
C. Pressing and holding either
button will cause the set-point to scroll rapidly in either direction.
Temperature
Set-point
Actual
Temperature
Increase
Temperature
Set-point
Decrease
Temperature
Set-point
Decon Cycle
Active Indicator
Switch to
initiate Decon