If the incubator temperature display reads 40.0
C and the calibrated independent sensor
shows 40.3
C, set the i.CAL offset value to 0.3
C. If the calibrated independent sensor
shows 39.6
C, then the entered offset should be negative. In this example the required
offset to i.CAL would be -0.4
C.