If humidity calibration is needed, the following steps can be taken:

Locate the reference instrument’s temperature sensor in close proximity to the

cabinet’s geometric center. Be sure that the stabilization times described earlier have

been satisfied prior to performing this calibration. Press the advance key until the

green display reads i.CAL (calibrate). Pressing the UP arrow will increase the Humidity

calibration offset by 1%. Pressing the down arrow will decrease the Humidity

calibration offset by 1%. Pressing and holding either button will rapidly scroll the

calibration offset. When finished, press the infinity key to return to the main menu.

Humidity calibration example

If the chamber temperature display reads 80% and the calibrated independent sensor

shows 83%, set the i.CAL offset value to 3.0%. If the calibrated independent sensor

shows 74

C, then the entered offset should be negative. In this example the required

offset to i.CAL would be -6.0%.