6022, 6023, 6041, 6042 & 6043 have one controller each. Chambers 6020, 6025,

6040 & 6045 have two controllers each.

Connect shielded wires to the RS-485 Communications terminal blocks.

Communications wires should be shielded and routed away from power wires.

Maximum distance of total wire is 2000 feet. When connecting multiple chambers,

route wires in a daisy chain fashion. A termination resistor of 120 ohms may be placed

across RS-485(-) and RS-485(+) of the last chamber in the daisy chain.

For PC’s having an RS-232 serial port (9-pin D-Sub) connection, an RS-485 to RS-232

converter may be used. Recommended sources are B&B Electronics (P/N 4850I9TB)

and CMC-Connecticut Micro Computer (ADA485L).

SFTW301 software is available from CARON to load onto a PC to monitor and record

data. This data storage is not 21CFR Part 11 compliant.