SECTION 5 – OPERATION
With the chamber properly installed and the appropriate utilities connected, the power
switch on the lower right of the control bezel can be turned on. Within a few minutes,
the temperature and humidity will begin to approach set-points. Allow the unit to
stabilize for 12 hours before use or prior to making any calibration adjustments.
Changing the Temperature Set-point
To set the temperature set-point, press the UP arrow to increase the temperature set-
point by 0.1
C. Press the DOWN arrow to decrease the temperature set-point by
0.1
C. Pressing and holding either button will cause the set-point to scroll rapidly in
either direction.
To turn off the temperature control system, but still display the chamber temperature,
press the EZ button. The words “OFF” will display in the set-point area. To toggle
back to controlled temperature, press the EZ button again.
If the unit is equipped with the deluxe controller package (DLUX301 or DLUX302),
additional ramp and soak features are available. Refer to Section 7 of the manual to
program these features.