Caron 6047 User Manual
Page 15
6020 & 6040 Series Operations Manual
Rev E 2/14/2012
Page 15 of 66
Connect each chamber to a grounded circuit.
Failure to do so could result in electrical shock.
Use only distilled or deionized water with a
resistivity between 50K
-CM and 1M
-CM and
a pH of greater than 6.5. Using water outside
this range will void your warranty.
Do not use water that contains chloramines.
Chloramines can damage internal rubber
gaskets resulting in leaks.
Connecting the Water Supply
To ensure proper operation, distilled or deionized water is required as a supply on units
that have humidity control. If these water sources are not available contact CARON
customer service.
A water inlet fitting on the back of the unit and ¼” black tubing are provided to connect
the water supply to the chamber. Connect an appropriate water supply to the fitting.
Incoming line pressure should be regulated to not exceed 80 psi.
If a Condensate Recirculator water recycling system was purchased as a water supply,
refer to its user’s manual for proper installation of the water supply.
Connecting Electrical Power
The unit requires a dedicated electrical outlet. See table below for model specific
power required and connection.
Model #
Power Requirements
Plug Connection
-1
115V, 60Hz, 16A FLA
NEMA 5-20
-2
230V, 60Hz, 10A FLA
NEMA 6-15
-3
230V, 50Hz, 8A FLA
CEE 7/7
When the required electrical connection is available, plug the provided power cord into
the unit and the electrical outlet.