6320 & 6340 Series Operations Manual

Rev G 9/16/2013

Page 31 of 61

Changing the CO

2

Set-point (CO2C301)

If an alternative CO

2

set-point is required, the following steps can be taken:

Press the up arrow to increase the CO

2

set-point by 0.1% CO

2

. Press the down arrow

to decrease the CO

2

set-point by 0.1% CO

2

. Pressing and holding either button will

cause the set-point to scroll rapidly in either direction.

If CO

2

is not required in the chamber, the CO

2

system can be disabled by turning off

the Enable CO

2

switch. This will disable the entire CO

2

system including the controller.

The system can be turned back on at anytime.

Operation of the Delux Controller System (DLUX301 & DLUX302)

The chambers can be purchased with upgraded controllers. The controllers have

additional features of RS485 communications, analog outputs, and ramp & soak.

RS485 Communications:

Each controller on the network (connected to the same PC or master device) must be

programmed with a unique address. If multiple chambers are being connected, each

controller will have to be assigned a unique address. Factory defaults for the

controllers are addresses 1 for temperature and 2 for humidity. As an example, a

second chamber would need to be assigned address 3 for temperature and 4 for

humidity. To assign new addresses to controllers, the controllers must be unlocked.

Refer to Section 13, Unlocking the Controllers.

CO

2

Set-

point

Increase CO

2

Set-point

Decrease

CO

2

Set-point

Enable CO

2

Actual CO

2

This manual is related to the following products: