6320 & 6340 Series Operations Manual

Rev G 9/16/2013

Page 20 of 61

Connecting Communications or Analog Outputs (DLUX301 & DLUX302)

With the purchase of DLUX301 or DLUX302, the controllers are upgraded with

additional features such as RS485 communications, and analog outputs. A set of

terminals are provided to connect to RS485 communications and analog outputs.

Analog Outputs

DLUX301 & DLUX302 provide analog outputs as either milliamps (0-20mA, 4-20mA) or

voltage (0-5V, 1-5V, 0-10V, 0-20V) signal output that represents each of the displayed

temperature (and humidity) values. These options can be used for connection to in-

house data acquisition, recorder, or alarm system.

Factory default settings are as follows:

Parameter

Analog Output

Corresponding Value

Temperature

0 – 5 V

0 – 100 °C

Humidity

0 – 5 V

0 – 100 %rh

Connect shielded wires to the appropriate signal terminals: I(+) for current (mA) or V(+)

for voltage (DC). For both current and voltage outputs, COM(-) is common terminal.

The controller itself must be programmed for the voltage or current signal and

corresponding scale (see Operations Section)

Communications

RS-485 communications are intended to communicate with a PC using ModBus RTU

or Standard Bus. The maximum number of chambers connected to a single PC is

limited to 247 controllers (Modbus) or 16 controllers (Standard Bus). Chambers 6021,