6320 & 6340 Series Operations Manual
Rev G 9/16/2013
Page 22 of 61
Use only distilled or deionized water with a
resistivity between 50K
-CM and 1M
-CM and
a pH of greater than 6.5. Using water outside
this range will void your warranty.
Do not use water that contains chloramines.
Chloramines can damage internal rubber
gaskets resulting in leaks.
Connecting the Water Supply (HUMD301)
To ensure proper operation, distilled or deionized water is required as a supply on units
that have humidity control. If these water sources are not available contact CARON
customer service.
A water inlet fitting on the back of the unit and ¼” black tubing are provided to connect
the water supply to the chamber. Connect an appropriate water supply to the fitting.
Incoming line pressure should be regulated to not exceed 80 psi.
If a Condensate Recirculator water recycling system was purchased as a water supply,
refer to its user’s manual for proper installation of the water supply.
Tubing to water
inlet