6320 & 6340 Series Operations Manual

Rev G 9/16/2013

Page 19 of 61

High concentrations of carbon dioxide can

cause asphyxiation. The use of CO

2

monitors

and alarms is recommended for areas where

CO

2

can collect.

The CO

2

gas supply should be 99% pure and

should not contain a siphon tube. Gas pressure

to the unit must be regulated to 15-20PSI.

Failure to do so could cause tubing to burst.

Connecting a CO

2

supply (CO2C301)

The CO

2

supply should be 99% and not have siphon tubes. CO

2

pressure should be

regulated to 15-20 psi. CO

2

tank regulators can be purchased through CARON

customer service. Once the cylinder regulator is installed, connect the outlet of the

regulator to the hose barb fitting using the tubing and clamps provided. An inline

HEPA filter is provided to remove any contaminants in the CO2 gas supply. Check the

connections closely for leaks.

If the unit is equipped with a built in gas guard system, there will be 2 gas inlets. Each

of the inlets should be connected to an individual gas tank as described above.