Controlling the Fluorescent & LED Lighting

The plant growth chamber comes with a fluorescent lighting system used to simulate

day and night testing. There are separate temperature and humidity (optional) set-

points that correspond with the lights on (day) and lights off (night). The set-points can

also be made the same if continuous conditions are needed throughout the light

cycles.

Each fluorescent light bank has two lamps and is mounted on the underside of a shelf.

The banks can be located on the same or separate shelves. Each light bank has a

corresponding switch that enables that light bank to come on when the timer is on.

Only the light banks that have their individual switch turned on will illuminate. The

green indicator light above the light bank switch signifies when each light bank is on.

The number of light banks varies depending upon the chamber model. The LED’s

(models 6323 & 6343) are located on the side walls. For units equipped with ramping

controllers (DLUX301 & DLUX302), the timer can be used to trigger an event in a

profile to switch between two set-points (if programmed accordingly).

Models 6322, 6323, 6342 & 6343

Individual

Light Bank

switches

Light Power

Switch

Light Bank

Indicators

Light Cycle

Timer

Models 6320, 6321 & 6340

Model 6341