Replacing Fluorescent Light Lamps

The fluorescent light lamps have a maximum recommended operating life of 5000

hours. It is recommended that the lamps be replaced to ensure proper internal light

intensity.

When the lamps have reached 4250 operating hours, the ‘Replace Lamps’ light on the

Alarm control panel will illuminate continuously. This is a warning that the lamps need

replaced soon.

Once the lamps have reached 5000 hours, the ‘Replace Lamps’ light will flash. This

indicates that to insure proper light levels, the lamps need replaced immediately.

Regardless of the silence switch position, no audible alarm will sound.

Once the light lamps have been replaced, reset the alarm by pressing the ‘Reset’

button. The button is recessed so use a pen or something similar. Once reset, the

alarm ‘Replace Lamp’ light will turn off.

To order replacement lamps, contact CARON’s service department at 740-373-6809 or

[email protected].