7.7 Power Settings

The power limit settings (parameter

OP.Hi

) for the STF models on various supply voltages are as

follows. The figures represent the maximum percentage of time that controlled power is supplied
to the elements. Do not attempt to “improve performance” by setting a value higher than the
correct one from the table.

Single phase

2-phase+N

3-phase delta

Model

Volts: 220V

230V

240V

380V

400V

415V

220V

230V

240V

STF 14/--/180

-

45

41

-

-

-

-

-

-

STF 14/--/450

74

68

62

53

48

44

-

-

-

STF 14/--/610

71

65

60

71

65

60

-

-

-

STF 15/--/180

49

45

41

-

-

-

-

-

-

STF 15/--/450

64

59

54

64

59

54

-

-

-

STF 15/--/450 *

82

75

69

59

54

49

STF 15/--/610

78

71

65

78

71

65

-

-

-

STF 16/--/180

38

35

32

-

-

-

-

-

-

STF 16/--/450

51

47

43

51

47

43

30

28

25

STF 16/--/610

63

57

53

63

57

53

28

29

27

* 5000W model up to late 2000.

8.0 S

PECIFICATIONS

Carbolite reserves the right to change specifications without notice

.

8.1 Models Covered by this Manual

MODEL

Max.

Temp.

Max.

Power

Work

Tube

Bore

1

Work

Tube

Length

Heated
Length

Net

Weight

2

(°C)

(kW)

(mm)

(mm)

(mm)

(kg)

Tube furnaces heated by Silicon Carbide elements
STF 14/--/180

1400°C

1.1

25-50

600-900

180

22

STF 14/--/450

1400°C

4.5

25-75

900-1200

450

35

STF 14/--/610

1400°C

5.5

25-75

1200-1500

610

40

STF 15/--/180

1500°C

1.5

25-50

600-900

180

22

STF 15/--/450

1500°C

5.5

25-75

900-1200

450

35

STF 15/--/610

1500°C

6.0

25-75

1200-1500

610

40

STF 16/--/180

1600°C

2.5

25-75

600-900

180

22

STF 16/--/450

1600°C

6.0

25-75

900-1200

450

35

STF 16/--/610

1600°C

7.0

25-75

1200-1500

610

40

note 1:

Standard tube internal diameters of 25, 50 and 75 are available;

intermediate tube sizes can also be fitted.

note 2:

Weights are approximate for horizontal models and do not include fittings

or vertical stands.