beautypg.com

Circuit diagrams – Carbolite STF Series User Manual

Page 14

background image

STF 14-16


14

LMF08 – 1.13

7.0 C

IRCUIT

D

IAGRAMS

EMC Filters (if fitted): dependent on the model there may be one filter, or more than one fitted in
parallel. The circuit diagram examples do not show multiple filter arrangements.

7.1 Single Phase

Applies for live-to-live supplies as well as live-to-neutral

7.2 2-phase with neutral

As the following 3-phase diagram, except that one phase is not present.

Note that if the phase current is less than 25A/phase, a direct safety switches may be fitted in the
element circuit, rather than the switch shown in the coil circuit.

7.3 3-phase with neutral

e

l

e

m

e

n

t

(s)

e

l

e

m

e

n

t

(s)

F3

F1

N

F2

L

PE

e

l

e

m

e

n

t

(s)

Instrument Switch

coil

temperature

controller

thermocouple

overtemp.

controller

thermocouple

SSR

heat

on

F3

SSR

SSR

heat

on

F3

heat

on

F3

Filter (if fitted)

if over-

temperature

fitted

contactor

if fitted: 2- or 3- circuit

models

Heater Switch

e

l

e

m

e
n

t

(s)

e

l

e

m

e
n

t

(s)

F1

N

L2

L3

F2

L1

PE

e

l

e

m

e
n

t

(s)

Instrument Switch

coil

temperature

controller

thermocouple

overtemp.

controller

thermocouple

SSR

heat

on

F3

SSR

SSR

heat

on

F3

heat

on

F3

Filter (if fitted)

contactor

note on (2- and) 3-phase: depending on filter(s) fitted, there may be (2

or) 3 separate neutral wires from the elements to the neutral supply.

if over-

temperature

fitted

Heater Switch