Fuses & power settings, On 8.1 – Carbolite TVS Series User Manual

Page 15

background image

HZS, TVS


MF24 – 3.44

15

8.0 F

USES

&

P

OWER

S

ETTINGS

8.1 Fuses

F1

Internal
supply
fuses

Fitted if supply cable fitted.
Fitted on board to some types of EMC filter.

on-board and up to 16 Amps:
38mm x 10mm type F
other: GEC Safeclip

F2

Auxiliary
circuit
fuses

Fitted on board to some types of EMC filter.
May be omitted up to 25Amp/phase supply
rating.

2 Amps glass type F
On board: 20mm x 5mm
Other: 32mm x 6mm

F3

Heat Light
fuses

May be omitted up to 25 Amp/phase supply
rating.

2 Amps glass type F
32mm x 6mm

Customer
fuses

Required if no supply cable fitted.
Recommended if cable fitted.

See rating label for amperage;
see table below for fuse rating.

Access to internal fuses is by removal of the back panel of the control box.

Model
(from March 2001)

phases

Volts

Supply Fuse

Rating

HZS & TVS 12/600

1-phase

200-240

15-16A **

HZS & TVS 12/600

3-phase + N

380/220-415/240

10A

HZS & TVS 12/600

3-phase delta

200-240

10A

HZS & TVS 12/600E

1-phase

200-240

15-16A **

HZS & TVS 12/600E

3-phase + N

380/220-415/240

5A

HZS & TVS 12/600E

3-phase delta

200-240

10A

HZS & TVS 12/900

1-phase

200-240

25A

HZS & TVS 12/900

3-phase + N

380/220-415/240

10A

HZS & TVS 12/900

3-phase delta

200-240

15A

HZS & TVS 12/900E

1-phase

200-240

25A

HZS & TVS 12/900E

3-phase + N

380/220-415/240

10A

HZS & TVS 12/900E

3-phase delta

200-240

12.5A

Other models or voltages: check the rating label for details of the supply.

Before March 2001 – please check with Carbolite if in doubt

8.2 Power Settings

The standard power limit settings (parameter

OP.Hi

) are as follows:

Volts:

HZS or TVS 12/--

100V
200V


208V

110V
220V
380V


230V
400V

120V
240V
415V

127V
254V
440V

before March 2001 – all models

100

100

100

100

100

89

from March 2001 **

100

100

89

82

75

67

See section 3.7 for further detail.

** Exceptions: models with 240V elements may be supplied in the following circumstances; they
have power settings of 100%.

- furnaces without controls - model /600 for use on a UK 13A supply