Fuses & power settings, On 8.1 – Carbolite TVS Series User Manual
Page 15
HZS, TVS
MF24 – 3.44
15
8.0 F
USES
&
P
OWER
S
ETTINGS
8.1 Fuses
F1
Internal
supply
fuses
Fitted if supply cable fitted.
Fitted on board to some types of EMC filter.
on-board and up to 16 Amps:
38mm x 10mm type F
other: GEC Safeclip
F2
Auxiliary
circuit
fuses
Fitted on board to some types of EMC filter.
May be omitted up to 25Amp/phase supply
rating.
2 Amps glass type F
On board: 20mm x 5mm
Other: 32mm x 6mm
F3
Heat Light
fuses
May be omitted up to 25 Amp/phase supply
rating.
2 Amps glass type F
32mm x 6mm
Customer
fuses
Required if no supply cable fitted.
Recommended if cable fitted.
See rating label for amperage;
see table below for fuse rating.
Access to internal fuses is by removal of the back panel of the control box.
Model
(from March 2001)
phases
Volts
Supply Fuse
Rating
HZS & TVS 12/600
1-phase
200-240
15-16A **
HZS & TVS 12/600
3-phase + N
380/220-415/240
10A
HZS & TVS 12/600
3-phase delta
200-240
10A
HZS & TVS 12/600E
1-phase
200-240
15-16A **
HZS & TVS 12/600E
3-phase + N
380/220-415/240
5A
HZS & TVS 12/600E
3-phase delta
200-240
10A
HZS & TVS 12/900
1-phase
200-240
25A
HZS & TVS 12/900
3-phase + N
380/220-415/240
10A
HZS & TVS 12/900
3-phase delta
200-240
15A
HZS & TVS 12/900E
1-phase
200-240
25A
HZS & TVS 12/900E
3-phase + N
380/220-415/240
10A
HZS & TVS 12/900E
3-phase delta
200-240
12.5A
Other models or voltages: check the rating label for details of the supply.
Before March 2001 – please check with Carbolite if in doubt
8.2 Power Settings
The standard power limit settings (parameter
OP.Hi
) are as follows:
Volts:
HZS or TVS 12/--
100V
200V
208V
110V
220V
380V
230V
400V
120V
240V
415V
127V
254V
440V
before March 2001 – all models
100
100
100
100
100
89
from March 2001 **
100
100
89
82
75
67
See section 3.7 for further detail.
** Exceptions: models with 240V elements may be supplied in the following circumstances; they
have power settings of 100%.
- furnaces without controls - model /600 for use on a UK 13A supply