beautypg.com

Carbolite CR Series User Manual

Page 8

background image

Clean Room Ovens


8 MF21 – 3.07

The new solid state relay contains a built-in MOV which protects it from short periods of excess
voltage. If the old relay had a separate disc-shaped "MOV" connected between the high voltage
terminals of the old relay, discard the old MOV.

Replace the removed panel.

5.6 Thermocouple Replacement

Remove the appropriate cover or control panel and the internal cover as given above.

Make a note of the thermocouple connections, and how the thermocouple is placed and fixed.
Colour codings are:

negative

positive (type K)

white

green

Disconnect the thermocouple from its controller terminals or terminal block.

Re-assemble with the new thermocouple observing the colour coding.

5.7 Element Replacement

Remove the control panel and the internal cover as given above, and the rear motor cover.

Disconnect the wires from the element terminals. Remove any starlock washers - these may need
to be cut with wire cutters. Remove any clips holding the element inside the chamber, and
withdraw the element.

Reverse the procedure with the new element.

Run the oven at a low temperature and check that it is controlling properly, to find out whether the
element failure was caused by a fault in the control circuit.

5.8 Fuse Replacement

Models up to 180 litre: remove the front control panel. Larger models: remove the top or side
panel of the control compartment.

Depending on the model, supply fuses and control circuit fuses may be mounted in their own
holders, or may be on a circuit board which contains an EMC filter. The fuses are marked with
their ratings.

Take care not to disconnect the wires leading from the EMC filter without first recording their
positions: they must be reconnected to the correct terminals.