3.2 Heater/Stirrer Set-Up and Programming
3.2a General Information
Circulator Pump
Pump speed is selected via the Main Menu.
The MINIMUM setting is used for DGGE applications.
MAXIMUM is recommended where temperature varies frequently and there is a need for fast recovery or when
pumping to multiple external units.
Speed
Maximum Pump Outlet Ratings
Maximum Pump Outlet Ratings
Selection
Line Frequency = 60Hz
Line Frequency = 50Hz
MAXIMUM
30 LPM / 5.0 PSI
22 LPM / 3.4 PSI
HIGH
24 LPM / 2.5 PSI
20 LPM / 1.7 PSI
MEDIUM
20 LPM / 1.7 PSI
17.5 LPM / 1.2 PSI
LOW
18 LPM / 1.2 PSI
15.8 LPM / 0.8 PSI
MINIMUM
15.7 LPM / 0.8 PSI
15 LPM / 0.7 PSI
The data in the table above are based on the following criteria:
a. Maximum pump outlet flow rate is measured in liters per minute (LPM) with no restriction on the pump outlet.
b. Maximum pump outlet pressure is measured in pounds per square inch (PSI) at no flow.
c. Water was used as the circulation fluid. Water has a viscosity of one centistoke. High viscosity or low-density fluids will change
these figures.
Filling the Reservoir
A liquid level that fully covers the heater coil, pump, over-temperature sensor, and at least one inch
(25mm) of the temperature sensor must be maintained. For heating applications, the TTGE tank
must
have at least 23 liters of buffer. When running the system with two dual cassettes, buffer volume should
be 24 liters.
minimum fill level
Temperature Sensor
Model Type
Temperature
Range
Reservoir
Capacity
Amps @
120V, 60Hz
Amps @
240V, 50Hz
Heat Only
Immersion Circulator
Ambient +5°C to
200°C*
N/A
11A
9.7A