Calibration

For most open air applications, the factory-set calibration should be correct.

Variations between the distance measured by the DCU-11 and the actual
distance are caused by environmental conditions such as temperature, humidity,
or chemical atmosphere. These environments can be compensated for by using
a calibration factor which calibrates the reading to match the actual distance.

MODE

DESCRIPTION

PARAMETERS

20

Distance Offset

Units = Mode 1
Range = 0-9999
Default = 0

Sets an offset for the display when measuring distance to a level.

MODE

DESCRIPTION

PARAMETERS

21

Offset Polarity

Range = NEG, POS
Default = NEG

Selects the direction of the offset. Adds or subtracts offset to reading.

MODE

DESCRIPTION

PARAMETERS

22

Calibration

Range = 0-9999

above the

Default = 1

Decimal Point

Sets the integer portion of the calibration factor.

MODE

DESCRIPTION

PARAMETERS

23

Calibration

Range = 0-9999

below the

Default = 0

Decimal Point

Sets the fractional portion of the calibration factor. (Decimal point is assumed