Mode Sheet (continued)
MODE DESCRIPTION PARAMETERS
EXPLANATION
10
Offset Distance
Range =-999 - +999
Default = 0
11
Temperature
Range = 0 - 2
Compensation
Default = 1
0 = OFF
1 = ON
2 = View/ON
12
Autosense
Range = 0 - 1
Default = 1
0 = OFF
1 = ON
13
Sensitivity
Range = 0 - 100%
Default = 100
14
Blanking
Units = Mode 1
Range = 0.5 to 16 ft.
Default = 1.00
15
Pulses
Units = Pulses
Range = 1 - 20
Default = 16
16
Software
Version
17
Reset
Range = 0 - 1
Default = 0
Sets an offset for the display.
Activates or deactivates internal
Temperature Compensation.
Activates or deactivates AutoSense.
Sets to top sensitivity level. Zero is
the lowest sensitivity settomg and
100 is the highest.
Sets a dead zone in front of the
transducer where echoes are
ignored.
Sets the number of ultrasonic
pulses transmitted in each burst. 20
being the strongest transmit setting.
Displays the Software Version
Resets the mode parameters to their
default values. Entering a 1 will
reset the parameters.